Leírás
RMCM-5 robotic mass comparator guarantees the utmost repeatability within 1 mg - 5 g range with the readability of 0.1 µg.
Function of Automatic Mass Standard Dissemination
RMCM-5 mass comparator facilitates a complete dissemination process. This is possible thanks to use of the intermediate mass standard magazine, which significantly shortens comparison duration and reduces wear and tear of the transport robot.
Real-Time Measurement of Ambient Conditions
The RMCM-5 mass comparator comes standard with a top-class thermo-hygro-barometer enabling real-time control of ambient conditions in three locations. The characteristic feature of the RMCM-5 is high readability of pressure, 0.001 hPa, humidity, 0.01 %, and temperature, 0.001 °C. Reliability of ambient conditions measurement carried out using the thermo-hygro-barometer is confirmed by a calibration certificate.
A Mass Comparator and a Robotic Transport System in One Compact Device
This combination results with:
reduction of air drafts and vibrations from a robotic system in the course of comparison,
minimised risk of human error,
easier maintaining of stable environmental conditions inside the weighing chamber (thanks to a compact device dimensions).
Optional Equipment
Additional equipment includes a video camera enabling remote preview of comparison process in real time (Internet access required), and RMCS software. The RMCS allows unrestricted monitoring of the comparison process.
Remote, real-time preview of comparison process in real time is possible thanks to a video camera.
Universal magazine insert shape
Mass standard magazine features 120 positions. The RMCM-5 enables comparison of weights of all shapes compliant with OIML recommendations, using just one universal insert.
Parameters
- Maximum capacity [Max]: 6,1 g
- E1 Calibration Range: 1 mg ÷ 5 g
- E2 Calibration Range: 1 mg ÷ 5 g
- F1 Calibration Range: 1 mg ÷ 5 g
- F2 Calibration Range: 1 mg ÷ 5 g
- Readability [d]: 0,1 µg
- Stabilization time: 30 s
- Adjustment: internal (automatic)
- Electric compensation range: 0 ÷ 6,1 g
- Magazine positions: 120
- Weighing pan dimensions: 24×50 mm
- Standard repeatability [5% Max]: 0,25 µg
- Standard repeatability [Max]: 0,4 µg
- Permissible repeatability: 0,6 µg
- Operating temperature: +15 ÷ +30 °C
- Relative humidity: 40% ÷ 60%
- Operating temperature change rate: ±0,5°C/12h (±0,3°C/4h)
- Relative humidity change rate: ±5%/12h (2%/4h)
- Communication interface: 2×USB-A, USB-C, HDMI, Ethernet, Wi-Fi, Hotspot
- Display: 10″ graphic colour touchscreen
- Controling device dimensions: 249×170×72 mm
- Power supply: 110 – 240 V AC 50/60 Hz
- E0 Calibration Range: 1 mg ÷ 5 g *
- Device dimensions: 1112×958×1570 mm